1. Position the arm in the horizontal position during leveling.

2. Using a spirit (bubble) level, check the plumb of the monitor with the level.

3. If adjustment is necessary, loosen the locking nut and turn the hex socket head screw to move the leveling rod forward or backward.

4. Tighten the locking nut.