If the table does not level correctly when the level button is pushed on the pendant, the table's potentiometers will need to be reset. First the table needs to be manually put in the level position. Using a level put each section in the level position. Once this step is complete, using the pendant push the level, on/off, and leg up buttons at the same time until an audible beep is heard. This resets the potentiometers. Test the table for level by pushing the level button on the pendant.