6702,3602,3502,6302 Surgical Tables. No Pendant Function.

Modified on Wed, 2 Jan, 2019 at 12:23 PM

This Bulletin involves a known condition with the 6702, 3602,3502 and 6302 surgical tables. 


SYMPTOMS

  1. No pendant operation 
  2. No battery indicator light 
  3. No emergency - activate emergency stop button for alarm.


Cause: All the symptoms listed above occur when the pendant cable is stretched out to far which causes damage to the pendant cable and relay box.


Troubleshooting:  


NOTE: Do not plug the pendant into any other table as damage will occur.


  1.  Check all electrical connection prior to replacing components. 
  2.  Inspect Pendant and Cable for physical damage.
  3.  Refer to the products service manual for proper electrical testing


Correction:

  1. Pendant cable replacement
  2. Relay box replacement or send the Relay box in for repair.






 

Was this article helpful?

That’s Great!

Thank you for your feedback

Sorry! We couldn't be helpful

Thank you for your feedback

Let us know how can we improve this article!

Select at least one of the reasons
CAPTCHA verification is required.

Feedback sent

We appreciate your effort and will try to fix the article